Acquiring Waveforms
3- 36
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Overview
Control elements and resources
To set acquisition modes (Cont.)
To select
real-time or
equivalent-
time sampling
To select real-time sampling, interpolated real-time sampling,
or equivalent-time sampling:
6.
Touch the Horiz button. Select the Acquisition tab from
the Horiz/Acq control window,
Or select Horizontal/Acquisition Setup from the
Horiz/Acq menu to display the Acquisition Mode control
window. Select the Acquisition tab.
7.
Select the sampling mode:
H
Equivalent Time uses both the real-time and
equivalent-time sampling as appropriate (see Table
3--2 on page 3--44).
H
Real Time Only limits the instrument to real-time
sampling. Real Time Only prevents the very fast
horizontal scale settings from being selected.
H
Interpolated Real Time limits the instrument to
real-time sampling. If the instrument cannot
accurately get enough samples for a complete
waveform, it will interpolate.
Note.
The instrument will use the interpolation method
selected in the display menu to fill in the missing record
points — either linear or sin(x)/x interpolation. The
default method is sine(x)/x. See
Interpolation
on
page 3--45 for a discussion of interpolation.
Summary of Contents for CSA7154
Page 4: ......
Page 22: ...Preface xviii CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 42: ...Installation 1 20 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 58: ...Incoming Inspection 1 36 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 78: ...Rear Panel I O Map 2 12 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 458: ...Glossary Glossary 18 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...