Appendix B: Automatic Measurements Supported
B- 8
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
The T1 values are vertical and horizontal values associated with the leftmost
crossing point. These areas are used to establish the following directions:
H
TCross1
mean
, the horizontal mean of the left crossing point at TCross
1
H
TCross1
sigma
, the horizontal standard deviation of the left crossing point
at TCross
1
H
TCross1
p-p
, the horizontal peak-to-peak deviation of the left crossing point
at TCross
1
H
PCross1
mean
, the vertical mean of the left crossing point at PCross
1
The T2 values are vertical and horizontal values associated with the rightmost
crossing point. These areas are used to establish the following directions:
H
TCross2
mean
, the horizontal mean of the right crossing point at TCross
2
H
TCross2
sigma
, the horizontal standard deviation of the right crossing point
at TCross
2
H
TCross2
p-p
, the horizontal peak-to-peak deviation of the right crossing point
at TCross
2
The duty cycle distortion (DCD) values are horizontal values associated with the
rightmost crossing point at 50% of the eye height. These areas are used to
establish the DCD
p-p
, the horizontal peak-to-peak deviation of the left crossing
point at half the height of the eye.
T1 Values
T2 Values
DCD Values
Summary of Contents for CSA7154
Page 4: ......
Page 22: ...Preface xviii CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 42: ...Installation 1 20 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 58: ...Incoming Inspection 1 36 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 78: ...Rear Panel I O Map 2 12 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 458: ...Glossary Glossary 18 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...