Data (Ch1) transition occurs
within
14.88 ns after the
clock violating the hold time
limit.
The instrument recognizes
the violation and triggers at
the clock edge.
Cursors measure the
setup/hold violation zone
which equals setup time +
hold time (30 ns).
Figure 3- 27: Triggering on a Setup/Hold time violation
Sequential Triggering
In applications that involve two or more signals, you may be able to use
sequential triggering to capture more complex events. Sequential triggering uses
the A (Main) trigger to arm the trigger system, and then uses the B (Delayed)
trigger to trigger the instrument if a specific condition is met. You can choose
one of two trigger conditions:
H
Trig After Time: After the A trigger arms the trigger system, the instrument
triggers on the next B-trigger event that occurs after the Trigger Delay Time.
You can set the trigger delay time with the keypad or the multipurpose
knobs.
H
Trigger on n
th
Event: After the A trigger arms the trigger system, the
instrument triggers on the n
th
B event. You can set the number of B events
with the keypad or the multipurpose knobs.
NOTE
. The traditional delayed trigger mode called “Runs After” is served by
Horizontal Delay. You can use horizontal delay to delay acquisition from any
trigger event, whether from the A (Main) trigger alone or from a sequential
trigger that uses both the A (Main) and B (Delayed) triggers. See
Triggering
with Horizontal Delay On
on page 3--116 for more information.
Summary of Contents for CSA7154
Page 4: ......
Page 22: ...Preface xviii CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 42: ...Installation 1 20 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 58: ...Incoming Inspection 1 36 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 78: ...Rear Panel I O Map 2 12 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 458: ...Glossary Glossary 18 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...