Incoming Inspection
1- 26
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Channel buttons
Figure 1- 8: Channel button location
5.
Set up the instrument:
H
Push the front-panel
AUTOSET
button. This sets the horizontal and
vertical scale and vertical offset for a usable display and sets the trigger
source to the channel that you are testing.
H
Touch the
Vert
button and then touch
Offset
. Confirm that the Ch1
Offset is 1.8 V (0.0 V if not using a probe).
6.
Verify that the channel is operational:
Confirm that the following statements
are true.
H
Verify that the vertical scale readout and the waveform amplitude for the
channel under test are as shown in Table 1--3.
Table 1- 3: Vertical settings
CSA7404, CSA7154, TDS7404, TDS7254,
TDS7154, TDS6604, & TDS6404
TDS7104 &
Setting
With P7240 or P7260
Without P7240
TDS7104 &
TDS7054
Scale
100 mV
500 mV
500 mV
Waveform amplitude 4 divisions
0.5 divisions
2 divisions