Measuring Waveforms
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
3- 181
Overview
Related control elements and resources
To compensate passive probes (Cont.)
For further
assistance
8.
Touch the Help button to access the online assistance.
See page 3--283 to learn about using online help.
You can adjust a relative time delay for each channel. This lets you align the
signals to compensate for signals that may come in from cables of differing
lengths. The instrument applies deskew values after it completes each acquisi-
tion; therefore, the deskew values do not affect logic triggering. Also, deskew
has no affect on fast acquisition XY and XYZ display formats.
CSA7000 Series, TDS7000 Series, and TDS6404: To deskew channels, use the
procedure that follows.
TDS6604: To deskew channels, use the procedure in the manual that came with
your deskew fixture.
Overview
To deskew channels
Related control elements and resources
Prerequisites
1.
Instrument should be powered on. Allow the instrument
and probes 20 minutes to warm up.
See page 3--34 for acquisition setup and
Power on the
Instrument
on page 1--9.
To Deskew Channels
Summary of Contents for CSA7154
Page 4: ......
Page 22: ...Preface xviii CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 42: ...Installation 1 20 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 58: ...Incoming Inspection 1 36 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 78: ...Rear Panel I O Map 2 12 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 458: ...Glossary Glossary 18 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...