Measuring Waveforms
3- 178
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
Overview
Related control elements and resources
To calibrate probes (Cont.)
Optimize gain
and offset
accuracy
2.
Connect the fixture to the instrument (see
To Connect
the Probe Calibration Fixture
on page 3--173).
3.
Remove the small jumper from the fixture.
4.
Connect one probe to the fixture.
5.
If compensating a passive probe, first perform the
To
Compensate Passive Probes
procedure on page 3--180.
6.
From the toolbar, touch the
VERT
button to display the
instrument Vertical setup control window.
7.
Touch the Probe
Cal
button to display the vertical Probe
Cal control window.
8.
Select the instrument channel to which the probe is
attached.
Note:
Clear Probecal Erase will erase any previous
calibration data for the attached probe.
9.
Touch the
Calibrate Probe
button.
10.
When the routine is finished, remove the connections
and replace the small jumper.