![Tektronix CSA7154 User Manual Download Page 53](http://html1.mh-extra.com/html/tektronix/csa7154/csa7154_user-manual_1077967053.webp)
Incoming Inspection
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
1- 31
CSA7404, CSA7154, TDS7404, TDS7254,
TDS7154, TDS6604, & TDS6404
BNC cable from PROBE
COMPENSATION output
to CH 1 input
TDS7104 & TDS7054
BNC cable from PROBE
COMPENSATION
output to CH 1 input
Figure 1- 10: Setup for trigger test
4.
CSA7404, CSA7154, TDS7404, TDS7254, TDS7154, TDS6604, &
TDS6404:
Touch the
Vert
button and then touch
Offset
. Adjust the Ch1 Offset to
0.8 V
using the multipurpose knob.
5.
CSA7404, CSA7154, TDS7404, TDS7254, TDS7154, TDS6604, &
TDS6404:
Set the Vertical
SCALE
to
100 mV
per division.
6.
Verify that the main trigger system operates:
Confirm that the following
statements are true.
H
The trigger level readout for the A (main) trigger system changes with
the trigger-
LEVEL
knob.
H
The trigger-
LEVEL
knob can trigger and untrigger the square-wave
signal as you rotate it. (Leave the signal
un
triggered).
H
Pushing the front-panel trigger
LEVEL
knob sets the trigger level to the
50% amplitude point of the signal and triggers the signal that you just
left untriggered. (Leave the signal triggered.)
Summary of Contents for CSA7154
Page 4: ......
Page 22: ...Preface xviii CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 42: ...Installation 1 20 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 58: ...Incoming Inspection 1 36 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 78: ...Rear Panel I O Map 2 12 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 458: ...Glossary Glossary 18 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...