Measuring Waveforms
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
3- 175
Overview
Related control elements and resources
To connect the probe calibration fixture (Cont.)
Connect the
probe
Warning.
To avoid personal injury, use care while connecting
probe tips to the square pins on the fixture. The ends of the
square pins are sharp.
Refer to these diagrams to attach your probe tip to the
fixture. Refer to the ground symbols on the fixture to
establish the correct polarity.
P6243 or P6245
Connect the probe tip to the short pin and the probe ground
to the long pin as shown.
Square-pin
adapter
Square-pin
ground socket
P7260
Connect the probe to the two long pins. Connect the probe
ground to the long pin that is marked with the ground symbol.
P6139A
Connect the probe tip and the ground lead to the two
terminals as shown.
B side
connection
A side
connection
Summary of Contents for CSA7154
Page 4: ......
Page 22: ...Preface xviii CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 42: ...Installation 1 20 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 58: ...Incoming Inspection 1 36 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 78: ...Rear Panel I O Map 2 12 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 458: ...Glossary Glossary 18 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...