Incoming Inspection
1- 32
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
7.
Verify that the delayed trigger system operates:
a.
Set up the delayed trigger:
H
From the
Trig
menu, select
A
→
B Sequence . . .
. This displays the
A
→
B Sequence tab of the trigger setup control window.
H
Click the
Trig After Time
button under A Then B.
H
Click the
B Trig Level
control in the control window.
b.
Confirm that the following statements are tru
e:
H
The trigger-level readout for the B trigger system changes as you
turn the lower multipurpose knob.
H
As you rotate the lower multipurpose knob, the square-wave
probe-compensation signal can become triggered and untriggered.
(Leave the signal triggered.)
c.
Verify the delayed trigger counter:
H
Double click the
Trig Delay
control to pop up a numeric keypad for
that control.
H
Click on the keypad to enter a trigger delay time of
1 second
, and
then click
Enter
.
H
Verify that the trigger
READY
indicator on the front panel flashes
about once every second as the waveform is updated on-screen.
8.
Remove the test hookup:
Disconnect the equipment from the channel input
and the Probe Compensation output.
Equipment
required
One BNC cable
One 1.44 Mbyte, 3.5 inch DOS-compatible formatted disk.
CSA7404, CSA7154, TDS7404, TDS7254, TDS7154, TDS6604, &
TDS6404:
One TCA-BNC TekConnect adapter
Prerequisites
None
Check File System
Summary of Contents for CSA7154
Page 4: ......
Page 22: ...Preface xviii CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 42: ...Installation 1 20 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 58: ...Incoming Inspection 1 36 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 78: ...Rear Panel I O Map 2 12 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 458: ...Glossary Glossary 18 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...