Measuring Waveforms
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
3- 173
Overview
Related control elements and resources
To compensate the instrument (Cont.)
For further
assistance
6.
Touch the Help button to access the online assistance.
See page 3--283 to learn about using online help.
CSA7000 Series, TDS7000 Series, and TDS6404: To compensate or calibrate
probes you must connect the Probe Calibration and Deskew Fixture to the
instrument; use the procedure that follows.
TDS6604: To compensate or calibrate probes you must connect the Probe
Calibration and Deskew Fixture to the instrument; use the procedure in the
manual that came with your deskew fixture.
Overview
To connect the probe calibration fixture
Related control elements and resources
Prerequisites
1.
Instrument should be powered on. Allow a 20 minute
warm up.
See page 3--34 for acquisition setup and
Power on the
Instrument
on page 1--9.
Overview
This fixture converts the PROBE COMPENSATION output
into a set of test points that give you a convenient way to:
H
Optimize the instrument gain and offset accuracy at the
probe tip (see
To Calibrate Probes
on page 3--177).
H
Compensate (low-frequency) a passive probe (see
To
Compensate Passive Probes
on page 3--180).
H
Compensate for timing differences (deskew) between
probes (see
To Deskew Channels
on page 3--181).
To Connect the Probe
Calibration Fixture
Summary of Contents for CSA7154
Page 4: ......
Page 22: ...Preface xviii CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 42: ...Installation 1 20 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 58: ...Incoming Inspection 1 36 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 78: ...Rear Panel I O Map 2 12 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...
Page 458: ...Glossary Glossary 18 CSA7000 Series TDS7000 Series TDS6000 Series Instruments User Manual ...