Electrical connection
Proline Promass F 500 EtherNet/IP
52
3. Fold open the terminal cover.
4. Push the cable through the cable entry . To ensure tight sealing, do not remove the
sealing ring from the cable entry.
5. Strip the cable and cable ends and connect to the RJ45 connector.
6. Connect the protective ground.
7. Plug in the RJ45 connector.
8. Firmly tighten the cable glands.
This concludes the EtherNet/IP connection process.
Connecting the supply voltage and additional inputs/outputs
1.
3.
2.
4.
4 x
7.
TX 20
TX 20
6.
10 (0.4)
22 mm
8.
24 mm
5.
A0033831
1. Push the cable through the cable entry . To ensure tight sealing, do not remove the
sealing ring from the cable entry.
2. Strip the cable and cable ends. In the case of stranded cables, fit ferrules.
3. Connect the protective ground.
4. Connect the cable in accordance with the terminal assignment .
Signal cable terminal assignment:
The device-specific terminal assignment is
documented on an adhesive label in the terminal cover.
Supply voltage terminal assignment:
Adhesive label in the terminal cover or
5. Firmly tighten the cable glands.
This concludes the cable connection process.
6. Close the terminal cover.
7. Close the housing cover.
L
WARNING
Housing degree of protection may be voided due to insufficient sealing of the housing.
‣
Screw in the screw without using any lubricant.