Proline Promass F 500 EtherNet/IP
Operation options
97
Star topology
4
2
1
3
5
5
5
A0032078
29
Options for remote operation via EtherNet/IP network: star topology
1
Automation system, e.g. "RSLogix" (Rockwell Automation)
2
Workstation for measuring device operation: with Custom Add-On Profile for "RSLogix 5000" (Rockwell
Automation) or with Electronic Data Sheet (EDS)
3
Computer with Web browser (e.g. Internet Explorer) for accessing the integrated device Web server or
computer with operating tool (e.g. FieldCare, DeviceCare) with COM DTM "CDI Communication TCP/IP"
4
Ethernet switch
5
Measuring device
Ring topology
The device is integrated via the terminal connection for signal transmission (output 1) and
the service interface (CDI-RJ45).