Commissioning
Proline Promass F 500 EtherNet/IP
168
Linear expansion coefficient
Square expansion coefficient
Parameter overview with brief description
Parameter
Prerequisite
Description
Selection / User
interface / User
entry
Factory setting
Corrected volume flow calculation
–
Select reference density for
calculating the corrected
volume flow.
• Fixed reference
density
• Calculated
reference density
• External reference
density
• Current input 1
• Current input 2
• Current input 3
Calculated reference
density
External reference density
In the
Corrected volume flow
calculation
parameter, the
External reference density
option is selected.
Shows external reference
density.
Floating point
number with sign
–
Fixed reference density
The
Fixed reference density
option is selected in the
Corrected volume flow
calculation
parameter
parameter.
Enter fixed value for reference
density.
Positive floating-
point number
1 kg/Nl
Reference temperature
The
Calculated reference
density
option is selected in
the
Corrected volume flow
calculation
parameter.
Enter reference temperature
for calculating the reference
density.
–273.15 to 99
999 °C Country-specific:
• +20 °C
• +68 °F
Linear expansion coefficient
The
Calculated reference
density
option is selected in
the
Corrected volume flow
calculation
parameter
parameter.
Enter linear, medium-specific
expansion coefficient for
calculating the reference
density.
Signed floating-point
number
0.0 1/K
Square expansion coefficient
The
Calculated reference
density
option is selected in
the
Corrected volume flow
calculation
parameter
parameter.
For media with a non-linear
expansion pattern: enter the
quadratic, medium-specific
expansion coefficient for
calculating the reference
density.
Signed floating-point
number
0.0 1/K²
*
Visibility depends on order options or device settings
10.6.3 Carrying out a sensor adjustment
The
Sensor adjustment
submenu contains parameters that pertain to the functionality of
the sensor.