Diagnostics and troubleshooting
Proline Promass F 500 EtherNet/IP
230
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
412 Processing download
Download active, please wait
0x2000204
Status signal
C
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
431 Trim 1 to n
Carry out trim
0x2000004
Status signal
C
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
437 Configuration incompatible
1. Restart device
2. Contact service
0x1000060
Status signal
F
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
438 Dataset
1. Check data set file
2. Check device configuration
3. Up- and download new
configuration
0x400006A
Status signal
M
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
441 Current output 1 to n
1. Check process
2. Check current output settings
• 0x8000099
• 0x80000B6
Status signal
S
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
442 Frequency output 1 to n
1. Check process
2. Check frequency output settings
• 0x800008A
• 0x8000122
Status signal
S
Diagnostic behavior
Warning