Technical data
Proline Promass F 500 EtherNet/IP
252
16.3 Input
Measured variable
Direct measured variables
• Mass flow
• Density
• Temperature
Calculated measured variables
• Volume flow
• Corrected volume flow
• Reference density
Measuring range
Measuring ranges for liquids
DN
Measuring range full scale values
min(F)
to
max(F)
[mm]
[in]
[kg/h]
[lb/min]
8
³⁄₈
0 to 2
000
0 to 73.50
15
½
0 to 6
500
0 to 238.9
25
1
0 to 18
000
0 to 661.5
40
1½
0 to 45
000
0 to 1
654
50
2
0 to 70
000
0 to 2
573
80
3
0 to 180
000
0 to 6
615
100
4
0 to 350
000
0 to 12
860
150
6
0 to 800
000
0 to 29
400
250
10
0 to 2
200
000
0 to 80
850
Measuring ranges for gases
The full scale values depend on the density of the gas and can be calculated with the
formula below:
max(G)
=
max(F)
· ρ
G
: x
max(G)
Maximum full scale value for gas [kg/h]
max(F)
Maximum full scale value for liquid [kg/h]
max(G)
<
max(F)
max(G)
can never be greater than
max(F)
ρ
G
Gas density in [kg/m³] at operating conditions
x
Constant dependent on nominal diameter
DN
x
[mm]
[in]
[kg/m
3
]
8
³⁄₈
60
15
½
80
25
1
90