Proline Promass F 500 EtherNet/IP
Diagnostics and troubleshooting
223
12.7 Adapting the diagnostic information
12.7.1 Adapting the diagnostic behavior
Each item of diagnostic information is assigned a specific diagnostic behavior at the
factory. The user can change this assignment for specific diagnostic information in the
Diagnostic behavior
submenu.
Expert → System → Diagnostic handling → Diagnostic behavior
You can assign the following options to the diagnostic number as the diagnostic behavior:
Options
Description
Alarm
The device stops measurement. The totalizers assume the defined alarm condition. A
diagnostic message is generated.
The background lighting changes to red.
Warning
The device continues to measure. The totalizers are not affected. A diagnostic message is
generated.
Logbook entry only
The device continues to measure. The diagnostic message is displayed only in the
Event
logbook
submenu (
Event list
submenu) and is not displayed in alternation with the
operational display.
Off
The diagnostic event is ignored, and no diagnostic message is generated or entered.
12.8 Overview of diagnostic information
The amount of diagnostic information and the number of measured variables affected
increase if the measuring device has one or more application packages.
In the case of some items of diagnostic information, the diagnostic behavior can be
changed. Change the diagnostic information → 223
12.8.1 Diagnostic of sensor
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
022 Temperature sensor defective
1. Check or replace sensor
electronic module (ISEM)
2. If available: Check connection
cable between sensor and
transmitter
3. Replace sensor
• 0x10000BE
• 0x10000BF
• 0x10000D5
• 0x10000D6
Status signal
F
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
046 Sensor limit exceeded
1. Inspect sensor
2. Check process condition
• 0x80000C8
• 0x80000CA
Status signal
S
Diagnostic behavior [from the
factory]
Alarm
1)