9.2
Overview of system files
System files
Version
Description
How to acquire
Electronic Data
Sheet (EDS
system file)
2.1
Certified in accordance with the
following ODVA guidelines:
• Conformance test
• Performance test
• PlugFest
Embedded EDS Support (File Object
0x37)
•
→ Download-
Area
• EDS system file integrated in the
device: can be downloaded via
the Web browser
Add-on Profile
Level 3
• Major
revision 1
• Minor
revision 1
System file for "Studio 5000"
software (Rockwell Automation)
→ Download-
Area
9.3
Integrating the measuring device in the system
A detailed description of how to integrate the device into an automation system (e.g.
from Rockwell Automation) is available as a separate document:
Select country → Automation → Digital communication → Fieldbus device
integration → EtherNet/IP
Protocol-specific data → 259.
9.4
Cyclic data transmission
Cyclic data transmission when using the device master file (GSD).
9.4.1
Block model
The block model shows which input and output data the measuring device makes available
for implicit messaging. Cyclical data exchange is performed using an EtherNet/IP scanner,
e.g. a distributed control system etc.
Measuring device
Control system
Transducer
Block
Input Assembly Fix
(Assem100) 44 byte
Permanently assigned
input group
→
EtherNet/IP
Mass flow fixed input
assembly
(Assem106) 32 byte
Permanently assigned
input group
→
Volume flow fixed input
assembly
(Assem107) 62 byte
Permanently assigned
input group
→
Concentration fixed input
assembly
(Assem109) 66 byte
Permanently assigned
input group
→
API Referenced corrections
fixed input assembly
(Assem110) 64 byte
Permanently assigned
input group
→
Water cut % fixed input
assembly
(Assem111) 80 byte
Permanently assigned