System integration
Proline Promass F 500 EtherNet/IP
108
Configuration 7: Exclusive Owner Multicast
Input Assembly Configurable
Instance
Size (byte)
Min. RPI (ms)
Input Assembly Configurable
Configuration
0 x 69
–
–
Output Assembly Fix
O → T Configuration
0 x 66
64
5
Input Assembly Fix
T → O Configuration
0 x 64
88
5
Configuration 8: Input Only Multicast
Input Assembly Configurable
Instance
Size (byte)
Min. RPI (ms)
Input Assembly Configurable
Configuration
0 x 69
–
–
Output Assembly Fix
O → T Configuration
0 x C7
–
–
Input Assembly Fix
T → O Configuration
0 x 65
88
5
Possible connections
No.
#1
#2
#3
#4
#5
#6
#7
#8
#9
Number of connections
1
1
1
1
1
1
1
1
1
Input assembly fixed (Assem100)
X
Mass flow fixed input assembly
(Assem106)
X
Volume flow fixed input assembly
(Assem107)
X
Input assembly custom (Assem101)
X
Viscosity fixed input assembly
(Assem108)
X
Concentration fixed input assembly
(Assem109)
X
API Referenced corrections fixed
input assembly (Assem110)
X
Water cut % fixed input assembly
(Assem111)
X
Heartbeat monitoring fixed input
assembly (Assem112)
X
Permanently assigned input group
Input assembly fixed (Assem100), 44 byte
Description
Byte
1.
File header (not visible)
1-4
2.
Current diagnosis
5-8
3.
Mass flow
9-12
4.
Volume flow
13-16
5.
Corrected volume flow
17-20
6.
Temperature
21-24
7.
Density
25-28