Diagnostics and troubleshooting
Proline Promass F 500 EtherNet/IP
214
LED
Color
Meaning
1
Supply voltage
Off
Supply voltage is off or too low.
Green
Supply voltage is ok.
2
Device status/module
status
Off
Firmware error
Green
Device status is ok.
Flashing green
Device is not configured.
Flashing red
A diagnostic event with "Warning" diagnostic
behavior has occurred.
Red
A diagnostic event with "Alarm" diagnostic behavior
has occurred.
Flashing red/green
The device restarts/self-test.
3
Network status
Off
The device does not have an EtherNet/IP address.
Green
EtherNet/IP connection is active.
Flashing green
The device has an Ethernet/IP address but no
EtherNet/IP connection is active.
Red
The EtherNet/IP address of the device has been
assigned twice.
Flashing red
EtherNet/IP connection is in the "time out" mode.
Flashing red/green
The device restarts/self-test.
4
Port 1 active:
Ethernet/IP
Off
Not connected or no connection established.
White
Connected and connection established.
Flashing white
Communication not active.
5
Port 2 active:
Ethernet/IP and
service interface (CDI)
Off
Not connected or no connection established.
Yellow
Connected and connection established.
Flashing yellow
Communication not active.
Proline 500
Different LEDs in the transmitter provide information on the device status.
1
2
3
4
5
A0029629
1
Supply voltage
2
Device status
3
network status
4
Port 1 active: EtherNet/IP
5
Port 2 active: EtherNet/IP and service interface (CDI)