Technical data
Proline Promass F 500 EtherNet/IP
278
c WLAN". The device acts as an Access Point and enables communication by
computer or a mobile handheld terminal.
Supported functions
Data exchange between the operating unit (such as a notebook for example) and the
measuring device:
– Upload the configuration from the measuring device (XML format, configuration
backup)
– Save the configuration to the measuring device (XML format, restore configuration)
– Export event list (.csv file)
– Export parameter settings (.csv file or PDF file, document the measuring point
configuration)
– Export the Heartbeat verification log (PDF file, only available with the "Heartbeat
Verification" application package)
– Flash firmware version for device firmware upgrade, for instance
– Download driver for system integration
– Visualize up to 1000 saved measured values (only available with the
Extended
HistoROM
application package → 283)
Webserver special documentation → 285
HistoROM
data management
The measuring device features HistoROM data management. HistoROM data management
comprises both the storage and import/export of key device and process data, making
operation and servicing far more reliable, secure and efficient.
When the device is delivered, the factory settings of the configuration data are stored
as a backup in the device memory. This memory can be overwritten with an updated
data record, for example after commissioning.
Additional information on the data storage concept
There are different types of data storage units in which device data are stored and used by the device:
Device memory
T-DAT
S-DAT
Available data
• Event logbook such as diagnostic events
for example
• Parameter data record backup
• Device firmware package
• Driver for system integration for exporting
via Web server, e.g.:
EDS for EtherNet/IP
• Measured value logging ("Extended
HistoROM" order option)
• Current parameter data record (used
by firmware at run time)
• Maximum indicators (min/max
values)
• Totalizer values
• Sensor data: nominal diameter etc.
• Serial number
• Calibration data
• Device configuration (e.g. SW options,
fixed I/O or multi I/O)
Storage location
Fixed on the user interface board in the
connection compartment
Attachable to the user interface board in
the connection compartment
In the sensor plug in the transmitter
neck part