Proline Promass F 500 EtherNet/IP
Diagnostics and troubleshooting
227
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
276 I/O module 1 to n faulty
1. Restart device
2. Change I/O module
• 0x100007B
• 0x1000081
Status signal
F
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
283 Memory content
1. Reset device
2. Contact service
• 0x10000E1
• 0x100016F
Status signal
F
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
302 Device verification active
Device verification active, please
wait.
0x20001EE
Status signal
C
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
311 Electronic failure
1. Do not reset device
2. Contact service
0x40000E2
Status signal
M
Diagnostic behavior
Warning
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
332 Writing in HistoROM backup failed
Replace user interface board
Ex d/XP: replace transmitter
0x10002C7
Status signal
F
Diagnostic behavior
Alarm
Diagnostic information
Remedy instructions
Coding of diagnostic
information (hex)
No.
Short text
361 I/O module 1 to n faulty
1. Restart device
2. Check electronic modules
3. Change I/O Modul or main
electronics
0x1000095
Status signal
F
Diagnostic behavior
Alarm