Commissioning
Proline Promass F 500 EtherNet/IP
128
Parameter
Description
Selection
Factory setting
Density unit
Select density unit.
Result
The selected unit applies for:
• Output
• Simulation process variable
• Density adjustment (
Expert
menu)
Unit choose list
Country-specific:
• kg/l
• lb/ft³
Reference density unit
Select reference density unit.
Unit choose list
Country-dependent
• kg/Nl
• lb/Sft³
Temperature unit
Select temperature unit.
Result
The selected unit applies for:
•
Electronic temperature
parameter
(6053)
•
Maximum value
parameter (6051)
•
Minimum value
parameter (6052)
•
External temperature
parameter (6080)
•
Maximum value
parameter (6108)
•
Minimum value
parameter (6109)
•
Carrier pipe temperature
parameter
(6027)
•
Maximum value
parameter (6029)
•
Minimum value
parameter (6030)
•
Reference temperature
parameter
(1816)
•
Temperature
parameter
Unit choose list
Country-specific:
• °C
• °F
Pressure unit
Select process pressure unit.
Result
The unit is taken from:
•
Pressure value
•
External pressure
parameter
• Pressure value
Unit choose list
Country-specific:
• bar a
• psi a
10.5.3 Configuring the communication interface
The
Communication
submenu guides you systematically through all the parameters that
have to be configured for selecting and setting the communication interface.
Navigation
"Setup" menu → Communication
‣
Communication
MAC address
Default network settings
DHCP client
IP address