Adjustments and Firmware Update
Adjustment Procedures
6-5
Figure
6-1, Calibration Password
4. Type your User Name Password and click on OK. The
Calibration Panel as shown in Figure 6-2 will appear.
Figure
6-2, Calibration Panel
NOTE
Initial factory adjustments require that the covers be
removed from the instrument. Field calibration does not
require re-adjustments of these factory settings unless the
unit was repaired in an authorized service center. Factory
adjustments are enclosed in parentheses to differentiate
from normal field calibration setups; bypass these
adjustments when performing field calibration.