TE5251
User Manual
5-6
Table
Chapter 5-3, Frequency Accuracy Using External 10 MHz Reference
5251 Setting
Error Limits
Counter Reading
Pass
Fail
10.000000000 MHz
±
1 Hz
50.000000000 MHz
±
5 Hz
Amplitude
Accuracy
Amplitude accuracy checks tests the accuracy of the output amplifier
and attenuators. Each channel has its own set of amplifiers and
attenuators and therefore, the accuracy is tested on each channel
separately. Amplitude path is checked for both the DAC route
(arbitrary and standard waveforms) and the DDS route (CW and
modulated waveforms).
Amplitude Accuracy,
DAC Output
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Termination:
50
Ω
feedthrough at the DMM input
Function: ACV
2. Connect 5251 Channel to the DMM input
3. Configure the 5251 as follows:
Frequency: 1
kHz
Output: On
Amplitude:
As specified in Table 4
Test Procedure:
1. Perform amplitude Accuracy tests using Table 4
Table
Chapter 5-4, Amplitude Accuracy, DAC output
5251 Amplitude
Setting Error
Limits DMM Reading
Pass
Fail
10.00V
3.534V,
±
60mV
5.000V
1.767V,
±
42mV
500mV
176.7mV,
±
6.7mV
50.00mV
17.67mV,
±
2.1mV
Amplitude Accuracy,
DDS Output
Equipment: DMM
Preparation:
1. Configure the DMM as follows:
Termination:
50
Ω
feedthrough at the DMM input
Function: ACV
2. Connect 5251 output to the DMM input
3. Configure the 5251 as follows:
Summary of Contents for 5251
Page 17: ...List of Figures continued xiv...
Page 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Page 37: ...TE5251 User Manual 2 2...
Page 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Page 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Page 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Page 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Page 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Page 290: ...TE5251 User Manual 2...