TE5251
User Manual
5-32
Test Results
Pass
Fail
4. Change sweep step to logarithmic and verify oscilloscope
exponential down waveform with properties as in 3 above
Test Results
Pass
Fail
Auxiliary
Counter/Timer
Operation
This tests the operation of the auxiliary counter/timer function. Note
that when you select the counter/timer function all other 5251
waveform generation are automatically purged and the instrument
operational mode is transformed to a stand-alone counter/timer.
Waveform generation is resumed as soon as the counter/timer
function is turned off.
Frequency
Equipment: Function Generator with at least 1 ppm accuracy
Preparation:
1. Configure the function generator as follows:
Frequency:
As required by the test
Wave:
Square
Amplitude 500
mV
2. Connect the function generator to the 5251 TRIG IN connector
3. Configure the 5251, as follows:
Auxiliary Function: Counter/Timer
Function: Frequency
Trigger Level:
0 V
Test Procedure:
1. Perform Frequency Measurement Accuracy tests using Table 18
Summary of Contents for 5251
Page 17: ...List of Figures continued xiv...
Page 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Page 37: ...TE5251 User Manual 2 2...
Page 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Page 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Page 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Page 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Page 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Page 290: ...TE5251 User Manual 2...