TE5251
User Manual
4-66
Run Mode
Commands
The Run Mode Commands group is used to synchronize device
actions with external events. These commands control the trigger
modes of the Model 5251. The generator can be placed in
Triggered, Gated or Burst mode. Trigger source is selectable from
an external source, an internal re-trigger generator or a software
trigger. Optional nodes were omitted from these commands. The
Run Mode settings affect all waveform shapes equally except when
using the modulated waveforms. In the case of modulated
waveform, the output idles on the carrier waveform until stimulated
to output a modulation cycle or burst of cycles. Factory defaults
after *RST are shown in bold typeface. Parameter low and high
limits are given where applicable
.
Table
4-7, Run Mode Commands
Keyword
Parameter Form
Default
:INITiate
[:IMMediately]
:CONTinuous
OFF | ON | 0 | 1
1
:TRIGger
[:IMMediate]
:BURSt
[:STATe]
OFF | ON | 0 | 1
0
:COUNt
1
to
1000000
1
:DELay
[:STATe]
OFF | ON | 0 | 1
0
:TIMe
200e-9
to
20
200e-9
:GATE
[:STATe]
OFF | ON | 0 | 1
0
:LEVel
-5 to 5
1.6
:SOURce
[:ADVance]
BUS | EXTernal | MIXed | TTLTrig 0-7 | STAR
EXT
:SLOPe
POSitive | NEGative
POS
:RETRigger
[:STATe]
OFF | ON | 0 | 1
0
:TIMe
200e-9
to
20
200e-9
:STAR
[:STATe]
OFF
|
ON
OFF
Summary of Contents for 5251
Page 17: ...List of Figures continued xiv...
Page 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Page 37: ...TE5251 User Manual 2 2...
Page 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Page 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Page 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Page 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Page 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Page 290: ...TE5251 User Manual 2...