TE5251
User Manual
5-26
Test Results
Pass
Fail
FM - Arbitrary
Waveforms
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
0.2 ms
Sampling Rate: 50 MS/s at least.
Trace A View:
Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude:
1 V/div
2. Connect the 5251 output to the oscilloscope input, channel 1
3. Connect the 5251 SYNC to the oscilloscope input, channel 2
4. Configure model 5251 controls as follows:
Waveform:
Modulated
Modulation:
FM
Mod Waveform: Arbitrary
Carrier Freq:
1 MHz
FM SCLK:
2.5 MS/s
Sync: On
Output: On
5. Using ArbConnection prepare, open the FM Composer and
download the following waveform:
Wavelength: 4000
points
Waveform:
4 cycles sinewave
Deviation: 0.5
MHz
Test Procedure:
1. Verify FM operation on the oscilloscope as follows:
Waveform: Sine
Frequency: 2.5
kHz
Max A:
1.25 MHz
Min A:
750 kHz
Test Results
Pass
Fail
AM
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
0.5 ms
Trigger source: Channel 2, positive slope
Amplitude:
1 V/div
2. Connect the 5251 output to the oscilloscope input, channel 1
3. Connect the 5251 SYNC to the oscilloscope input, channel 2
Summary of Contents for 5251
Page 17: ...List of Figures continued xiv...
Page 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Page 37: ...TE5251 User Manual 2 2...
Page 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Page 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Page 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Page 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Page 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Page 290: ...TE5251 User Manual 2...