5
Performance Checks
Test Procedures
5-25
Sync:
On
Output: On
Test Procedure:
1. Verify Gated FM – standard waveforms operation on the
oscilloscope as follows:
Waveform:
Gated sine waveforms
Sine Frequency: 10 kHz
Gated Period:
1 ms
Max A:
1.25 MHz
Min A:
750 kHz
Test Results
Pass
Fail
Re-triggered FM
Bursts - Standard
Waveforms
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
0.2 ms
Sampling Rate: 50 MS/s at least.
Trace A View:
Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude:
1 V/div
2. Connect the 5251 output to the oscilloscope input, channel 1
3. Connect the 5251 SYNC to the oscilloscope input, channel 2
4. Configure model 5251 controls as follows:
Waveform:
Modulated
Modulation:
FM
Modulation Run Mode: Burst
Burst Count:
5
Carrier Freq:
1 MHz
Mod Frequency: 10 kHz
Deviation: 500
kHz
Sync: On
Re-trigger: On
Re-trigger Delay: 200
μ
s
Output: On
Test Procedure:
1. Verify re-triggered FM burst – standard waveforms operation
on the oscilloscope as follows:
Waveform:
Repetitive burst of 5-cycle sine waveforms
Sine Frequency: 10 kHz
Re-trigger delay: 200
μ
s
Max A:
1.25 MHz
Min A:
750 kHz
Summary of Contents for 5251
Page 17: ...List of Figures continued xiv...
Page 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Page 37: ...TE5251 User Manual 2 2...
Page 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Page 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Page 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Page 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Page 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Page 290: ...TE5251 User Manual 2...