TE5251
User Manual
5-22
4. Configure model 5251 controls as follows:
Waveform:
Modulated
Modulation:
FM
Carrier Freq:
1 MHz
Mod Frequency: 10 kHz
Deviation: 500
kHz
Sync: On
Output: On
Test Procedure:
1. Verify FM operation on the oscilloscope as follows:
Waveform: Sine
Frequency: 10
kHz
Max A:
1.25 MHz
Min A:
750 kHz
Test Results
Pass
Fail
2. Modify 5251 modulating waveform to triangle, then square and
ramp and verify FM waveforms as selected
Test Results
Pass
Fail
3. Move 5251 marker position to 1.25MHz and verify marker
position
Test Results
Pass
Fail
Triggered FM -
Standard Waveforms
Equipment: Oscilloscope, function generator
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
0.2 ms
Sampling Rate: 50 MS/s at least.
Trace A View:
Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude:
1 V/div
2. Connect the 5251 output to the oscilloscope input, channel 1
3. Connect the 5251 SYNC output to the oscilloscope input,
channel 2
4. Configure the function generator as follows:
Frequency 1
kHz
Summary of Contents for 5251
Page 17: ...List of Figures continued xiv...
Page 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Page 37: ...TE5251 User Manual 2 2...
Page 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Page 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Page 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Page 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Page 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Page 290: ...TE5251 User Manual 2...