TE5251
User Manual
4-62
Response
The 5251 will return the present marker position.
3D:RASTer<3D_freq>(?)
Description
This command will set the sample clock frequency for the 3D modulation profiler. The 3D waveforms must
be created using an external utility and downloaded to the 3D memory before this function can be used.
Parameters
Name
Range
Type
Default
Description
<3D_freq>
1 to
2.5e6
Numeric
1e6
Programs the sample clock frequency of the 3D
modulating waveform in units of S/s.
Response
The 5251 will return the present sample clock of the 3D modulating waveform value. The returned value will
be in standard scientific format (for example: 100mHz would be returned as 100e-3 – positive numbers are
unsigned).