TE5251
User Manual
5-20
SYNC Qualifier - Bit
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
As required by the test
Amplitude:
2 V/div
2. Connect 5251 SYNC output to the oscilloscope input
3. Configure model 5251 as follows:
Waveform: Sine
Output: On
SYNC: On
Test Procedure:
1. Verify trace on the oscilloscope shows synchronization pulses
at 1
μ
s intervals
Test Results
Pass
Fail
SYNC Qualifier -
LCOM
Equipment: Oscilloscope
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
As required by the test
Amplitude:
2 V/div
2. Connect the 5251 output to the oscilloscope input (1)
3. Connect the 5251 SYNC output to the oscilloscope input (2)
4. Configure model 5251 channel as follows:
Waveform: Sine
Run Mode:
Burst
Burst Count:
10
Re-trigger On
Re-trig period:
10
μ
s
Output: On
Test Procedure:
1. From ArbConnection, click on the Manual Trigger and verify
that trace on the oscilloscope shows synchronization pulse
having 10
μ
s pulse width. Verify that the SYNC is high for the
duration of the burst.
Test Results
Pass
Fail
Summary of Contents for 5251
Page 17: ...List of Figures continued xiv...
Page 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Page 37: ...TE5251 User Manual 2 2...
Page 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Page 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Page 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Page 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Page 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Page 290: ...TE5251 User Manual 2...