TE5251
User Manual
5-34
Pulse Width
Equipment: Function Generator with at least 1 ppm accuracy
Preparation:
1. Configure the function generator as follows:
Frequency:
As required by the test
Wave:
Square
Duty Cycle:
As required by the test
Amplitude 500
mV
2. Connect the function generator to the 5251 TRIG IN connector
3. Configure the 5251, as follows:
Auxiliary Function: Counter/Timer
Function: Pulse
Width
Trigger Level:
0 V
Test Procedure:
1. Perform Pulse Width Accuracy tests using Table 20
Table
Chapter 5-20, Pulse Width Measurement Accuracy
Function Generator Setting
Error Limits
5251
Counter Reading
Pass
Fail
Frequency Duty
Cycle
10 kHz
50 %
50.00
μ
s
±
100 ns
100 kHz
50 %
5.000
μ
s
±
100 ns
1 MHz
70 %
700 ns
±
100 ns
2. Change the counter/timer slope to Negative
3. With the last function generator setting in Table 28, change the
function generator duty cycle to 70 %
4. Verify that the pulse width reading is 300 ns
±
100 ns
Test Results
Pass
Fail
Totalize, Gated
Equipment: Function Generator with at least 1 ppm accuracy
Preparation:
1. Configure the function generator as follows:
Frequency: 1
MHz
Wave:
Square
Amplitude 500
mV
Run Mode:
Burst
Burst Count:
100
2. Connect the function generator to the 5251 TRIG IN connector
3. Configure the 5251, as follows:
Auxiliary Function: Counter/Timer
Function: Totalize,
Gated
Trigger Level:
0 V
Summary of Contents for 5251
Page 17: ...List of Figures continued xiv...
Page 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Page 37: ...TE5251 User Manual 2 2...
Page 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Page 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Page 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Page 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Page 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Page 290: ...TE5251 User Manual 2...