TE5251
User Manual
5-24
Waveform:
Modulated
Modulation:
FM
Modulation Run Mode: Burst
Burst: 5
Carrier Freq:
1 MHz
Mod Frequency: 10 kHz
Deviation: 500
kHz
Sync: On
Output: On
Test Procedure:
1. Verify Burst FM – standard waveforms operation on the
oscilloscope as follows:
Waveform:
Burst of 5 Sine waveforms
Sine Frequency: 10 kHz
Burst Period:
1 ms
Max A:
1.25 MHz
Min A:
750 kHz
Test Results
Pass
Fail
Gated FM - Standard
Waveforms
Equipment: Oscilloscope, function generator
Preparation:
1. Configure the oscilloscope as follows:
Time Base:
0.2 ms
Sampling Rate: 50 MS/s at least.
Trace A View:
Jitter, Type: FREQ, CLK.
Trigger source: Channel 2, positive slope
Amplitude:
1 V/div
2. Connect the 5251 output to the oscilloscope input, channel 1
3. Connect the 5251 SYNC output to the oscilloscope input,
channel 2
4. Configure the function generator as follows:
Frequency 1
kHz
Run Mode:
Continuous
Waveform:
Squarewave.
Amplitude:
2 V
Offset: 1
V
5. Connect the function generator output connector to the 5251
TRIG IN connector
6. Configure model 5251 controls as follows:
Waveform:
Modulated
Modulation:
FM
Mod Run Mode: Gated
Carrier Freq:
1 MHz
Mod Frequency: 10 kHz
Deviation: 500
kHz
Summary of Contents for 5251
Page 17: ...List of Figures continued xiv...
Page 35: ...TE5251 User Manual 1 18 This page was intentionally left blank...
Page 37: ...TE5251 User Manual 2 2...
Page 134: ...TE5251 User Manual 3 82 Figure 3 62 Log File Example...
Page 222: ...TE5251 User Manual 4 88 This page was left intentionally blank...
Page 258: ...TE5251 User Manual 5 36 This page was intentionally left blank...
Page 288: ...TE5251 User Manual 6 30 This page was intentionally left blank...
Page 289: ...1 Appendices Appendix Title Page A Specifications A 1...
Page 290: ...TE5251 User Manual 2...