SF-8200 Operation Manual
76
/
76
6) The liquid is not detected
Possible Causes:
Indoor humidity is too large or too dry, leading to poor capacitance environment, affecting the liquid
level sensing sensitivity;
Lack of reagent;
Solution:
Humidity should be suitably adjusted. Meet the humidity requirements of the working environment
of the instrument is 30% -75%;
Add reagent;
7) Cuvette position sensor failure
Possible Causes:
Cuvettes run out
Cuvette stuck
Cuvette position sensor failure;
Solution:
Recharge the cuvette roll
Open the left cover of machine, check through the window, and clean the channels.
Quality service please contact the service department of SUEECSS or local agent.