SF-8200 Operation Manual
66
/
76
4 EQUIPMENT ADVANCED FEATURES OPTIONS
4.1 CHANNEL SETTING
This function is only displayed and operated under "Engineer" user privilege.
Enable or disable the channel to be tested during testing by setting the state of certain
channels on / off.
Clotting Method Channel 1 is the first test channel close to the waste basket. The optical
method channel 1 is the first test channel close to the clotting method.
4.2 REAGENT AREA POSITION CHECKING
Enable the "Disable reagent area position detection" function by checking the box.
Attention! After checking, the middle reagent area and active reagent area does not automatically
recognize the reagent bottle’s position. Therefore, please set the reagent hole manually.
Please refer to Chapter2 to set the reagent hole manually.
4.3 SETTINGS FRO DIPOSE THE CUVETTE WHICH LACK REAGENT
In the "Device" tab of the "Options" interface, enable the function "Do not throw cups without rea-
gents" in the optical test by checking the checkbox.
Attention! After checking, the instrument will wait for operation when there is a lack of reagent alarm
during operation of the instrument. This can result in some tests overtaking the temperature. There may
be incorrect test results for these tests.
4.4 SETTINGS FOR CALCULATE THE MEAN VALUE AUTOMATICALLY
In the "Device" tab of the "Settings" screen, enable the function "Mean value calculation automati-
cally" by checking the checkbox.
After checking, the result of the report is the mean value of multiple test results, no longer the result
of the last test result is reported.