SF-8200 Operation Manual
53
/
76
When all the tests are completed, as shown below. The "Calibration" interface displays the "Meas-
ured Values" of all the calibration points and click the "Save" button. Pop-up confirmation dialog box,
store calibration data;
The newly added calibration curve is applied as the current effective curve, as shown below. Select
the latest curve from the list and click the "Apply" button.
The newly added calibration curve, applied as a valid curve, changes from "green dot" to "red dot"
in front of the list, thus completing the creation of a new calibration curve.
4) manual input
When calibrator is used as sample test, the test result can be input by manual input. Create a new
calibration curve.
Create a calibration curve, select "Manual Input", enable the "Concentration" setting window,
fill in the calibration value;
Click the "Add" button, the calibration point data is automatically added to the list, if the con-
centration is not correct, click Modify;
Clotting cali
Clotting cali
Clotting cali
Clotting cali
Clotting cali
Clotting cali
Clotting cali
Clotting cali
①
②
①
②
③