SF-8200 Operation Manual
62
/
76
2) Result parameters
The test interface displays the "main unit" of the test result, which is added by setting the "result
unit". For details, refer to Chapter 7.
3) Re-dilute
The Dilution range is determined by setting the "Low Limit" and "High Limit". And choose the "dilu-
tion rate" at which the limit is reached for the dilution / concentration test. Specific value According to the
"main unit" choice, the setting value is also different.
Automatic Dilution:
Check the "Auto Dilution" check box, the system will automatically run ac-
cording to the results of heavy dilution test function;
Manual Dilution:
Without check the "Automatic Dilution" checkbox. In the "Test" screen, right-click
the sample item that requires a heavy dilution test and manually add the Dilution Test.
Attention! The automatic dilute function expands the report range of test results and in-
creases reagent and sample consumption. Choose carefully according to your needs.
Re-dilute