SF-8200 Operation Manual
32
/
76
1.2 SAMPLE LOADING
A
B
C
D
E
F
G
I
H
J
A. Sample bar test tube location number
The tip of the sample bar is No. 1, and the hole nearest to the handle is No. 10.
B. Sample Number
Barcode recognition can be automatically entered or manually entered.
C. Sample tube type
Divided into two types of standard tubes and micro tubes.
Standard tube: blood coagulation test dedicated vacuum blood collection tube (2ml or 3ml specifi-
cations)
;
Micro tube: 1.5ml trace tube.
Warning: When testing with the specified micro tube, the sample tube holder should be
used as it will affect the sensitivity of the sample needle and cause incorrect analysis re-
sults.
D. Emergency settings
Set a priority sample for emergency testing by checking the box.
E. Sample rack number
The sample bars are numbered from right to left, with sample bar 1 on the far right and sample bar
6 on the left.
F. Test items list
Show all the test items that the instrument is currently open.