SF-8200 Operation Manual
23
/
76
3 REAGENT PREPARATION
3.1 PREPARE REAGENTS
Prepare the required amount of clotting reagents, buffers, and special wash solutions as expected
Item
Reagent
Dosage/Test
SFT/Test
PT
PT Reagent
100μl
120μl
APTT
APTT Reagent
50μl
——
Calcium Chloride Solution
50μl
60μl
TT
TT Thrombin Reagent
100μl
120μl
FIB
FIB Imidazole Buffer
90μl
——
FIB Thrombin Reagent
50μl
60μl
D-Dimer
D-Dimer R1
210μl
——
D-Dimer R2
70μl
85μl
FDP
FDP R1
210μl
——
FDP R2
70μl
85μl
AT-
Ⅲ
AT-
Ⅲ
buffer
45μl
186μl
AT-
Ⅲ
Thrombin Reagent
155μl
AT-
Ⅲ
substrate
45μl
55μl
SFW cleaning liquid
Dosage is about 200ml per test
regarding seven items
Considering that not all of the reagents in the reagent bottle are available, there is a dead space in
the reagent bottle that cannot be handled by the syringe and an unavoidable loss of sample. The fol-
lowing additional dosages are provided for each reagent bottle.
specification
diameter
(
mm
)
height
(
mm
)
purpose
Additional amount
required
2ml
18.5
35
Quality control calibration
500μl
5ml
22
40
Intermediate reagent, start reagent
800μl
15ml
30
45
Buffer
1500μl
25ml
29
58
SFT cleaning liquid
1500μl