I SF-8200 SIMPLE INTRODUCTION
1 OVERVIEW
SF-8200 fully automated coagulation analyzer for coagulation and anticoagulation of human
plasma, fibrinolysis and anti-fibrinolytic function of the test, to quickly and accurately analyze a large
number of samples.
This instrument is for in vitro diagnosis only.
Figure 1.1 SF-8200 Fully automated coagulation analyzer
2 INSTALLATION
2.1 INSTALLATION STATEMENT
The SF-8200 Automated Blood Coagulation Tester has been thoroughly inspected at the factory
and carefully packed to prevent damage due to vibration during shipping.
After delivery, please double-check the packaging to confirm that the instrument is free from any ex-
ternal damage.
SF-8200 automated coagulation analyzer can only be Beijing Succeeder Technology Inc. profes-
sional engineers or by the Company authorized engineers to install at the same time the initial set.
After the instrument is installed and running, it should not be moved normally. If you really need to move,
you should empty the pipes and trash basket waste, and fixed XYZ sampling arm mechanism, robot
movement mechanism and trash basket. There is a groove on each side of the bottom of the instrument.
When transporting, put your hand here and raise it horizontally.
2.2 INSTALLATION ENVIRONMENT
1) Choose a stable level surface to avoid functional errors when installing.
2) Make sure the installation space is large enough for operation and maintenance. Place the computer
on the left side, wash tub and waste tub on the right side, and make sure that the bin on the right side is
completely removable.
3) The network power must have a protective earth terminal. Internal protective grounding terminal