EN
0408-600-705 Rev-AB
4
www.stryker.com
Instructions
WARNINGS:
• Perform recommended inspection and testing as indicated in the
Inspection
and Testing
section.
• ALWAYS operate the equipment within the specified environmental condition
values. See the
Specifications
section.
• This equipment is suitable to use in a professional healthcare facility
environment.
• ALWAYS wear the surgical helmet with a surgical toga or surgical hood to
obtain the required personal protection.
• DO NOT cut or puncture the fiber-optic cable.
• ALWAYS turn the light source power off before connecting or disconnecting the
fiber-optic cable to or from the source.
• ALWAYS hold the strain relief near the end of the fiber-optic cable when
connecting or disconnecting the cable from the light source.
CAUTIONS:
• DO NOT scratch the headlight lens.
• DO NOT stretch or kink the fiber-optic cable. Failure to comply will damage
the cable and reduce the light output permanently.
• If using a dictation recorder or MP3 player, ALWAYS make sure the device has
standard 3.5 mm port(s). Failure to comply may result in helmet or accessory
damage.
NOTE:
Helmet position will vary from person to person depending on head size
and personal comfort preferences.
To Don a Surgical Helmet
WARNING:
DO NOT connect the power cable plug to the power pack
before donning the helmet.
NOTE:
ALWAYS perform the following steps in sequence. Failure to comply may
cause the red power loss indicator to flash. See the
Troubleshooting
section.
1. Apply the comfort pads inside the helmet as required (figure 1).
2. Connect the helmet power cable to the appropriate power cord (figures 2 and
2a). See the
Accessories
section.
NOTES:
• To disconnect the power cord, pull the connector sleeve back before
separating the connector plug and receptacle (figure 2).
• If using a dictation power cord, connect the pink-band plug into the
microphone receptacle and the green-band plug into the speaker receptacle
(figure 2a). Only one connector plug may be used based on the recorder or
player receptacles available.
• If using a dictation recorder or MP3 player, connect a earphone to the helmet
audio jack (figure 2b as required). See the
Accessories
section.
3. Loosen the headband adjustment knob (figure 3).
CAUTION:
DO NOT pull to extend or push to retract the front flexible air duct.
Failure to comply may result in product damage.
4. Put the helmet on your head while extending the height adjustment strap for a
comfortable fit (figure 4).
5. Turn the headband knob to secure the helmet (figure 5).
NOTES:
• Make sure the helmet fits securely on top of your head.
• Make sure the face mask ties are accessible to allow easy removal of the
face mask when donning the hood or toga.
WARNING:
Keep your fingers away from the fan during operation.
6. Slide a fully charged power pack into the holster by aligning the grooves on
the power pack with the holster. The power pack will snap into place
(figure 6).
NOTES:
• Make sure the indicator lights momentarily flash and the fan turns on
automatically. See the
Troubleshooting
section if required.
• If using a helmet with a battery-powered headlight, install an extended life
power pack into the holster.
7. Hang the holster from your waistband and swivel the holster to the desired
orientation (figure 7).
8. Press the variable control buttons to adjust the fan speed for optimal air flow.
See the
Troubleshooting
section if required (figure 8).
9. Rotate the adjustable air nozzle tip for optimal air flow (figure 9).
NOTES:
• If using a dictation recorder or MP3 player, see the instructions for use
supplied with the device for volume adjustment, then place the earphone in
your ear (figure 2b).
• You are now ready to scrub for surgery and don a hood or toga. See the
instructions for use supplied with the hood or toga.
• If the face shield lens begins to fog during use, see the
Troubleshooting
section
.
To Use a Surgical Helmet with a Headlight
To Use a Fiber-optic Headlight
CAUTIONS:
• ALWAYS remove the protective cap from the plug of the ACMI fiber-optic cable
before installing the plug into the fiber-optic light source receptacle.
• ALWAYS put the protective cap onto the plug of the ACMI fiber-optic cable
after use and before storage.
NOTES:
• If using a fiber-optic headlight, turn the light beam on and off and adjust the
brightness at the light source. See the instructions for use supplied with the
light source.
• A clamp may be used to fasten the fiber-optic cable to the gown to support
the weight of the cable.
1. Remove the protective cap on the male plug of the ACMI fiber-optic cable and
install the plug into any compatible ACMI fiber-optic light source receptacle
(figure A).
2. Adjust the headlight beam direction with the headlight adjustment handle
(figure C).
To Use a Battery-powered Headlight
NOTE:
Using the battery-powered headlight will decrease the power pack capacity
by half. See the instructions for use supplied with the power pack for capacity
values.
1. Make sure a power pack is installed.
2. Press both variable air flow control buttons simultaneously until the light turns
on or off, as desired
(figure B).
3. Adjust the headlight beam direction with the headlight adjustment handle
(figure C).
To Doff a Surgical Helmet
1. Flex the tab back to release the power pack, then lift the power pack from the
holster (figure D).
2. Loosen the headband knob and remove the helmet from your head.
3. Remove and discard the comfort pads from the helmet. Clean the helmet and
cord. See the
Cleaning
section. To clean the power pack, see the instructions
for use supplied with the Flyte charger.
Inspection and Testing
WARNINGS:
• Only individuals trained and experienced in the maintenance of reusable
medical devices should inspect and test this equipment.
• DO NOT use any equipment if damage is apparent or the inspection criteria
are not met.
• DO NOT disassemble or service this equipment, unless otherwise specified.
NOTE:
For service, contact your Stryker sales representative or call Stryker
customer service. Outside the US, contact your nearest Stryker subsidiary.
INTERVAL
INSPECTION CRITERIA
ACTION
Before
Use
Inspect equipment for damage, wear, or
missing components.
If damage is
apparent, replace
the equipment.
Inspect power cord and fiber-optic cable for
cuts or punctures.
Inspect headlight lens for scratches.
Inspect the power pack housing for cracks.
Check fan motor for excessive noise.
NOTE:
If any component must be discarded, see the
Disposal/Recycle
section.
Summary of Contents for Flyte 0408-600-000
Page 5: ...0408 600 705 Rev AB EN www stryker com 5 Instructions Figures 1 3 2 4 2a 5 2b 6 ...
Page 6: ...EN 0408 600 705 Rev AB 6 www stryker com Instructions Figures continued 7 A 8 B 9 C D ...
Page 14: ...ES 0408 600 705 Rev AB 14 www stryker com Instrucciones Figuras 1 3 2 4 2a 5 2b 6 ...
Page 15: ...0408 600 705 Rev AB ES www stryker com 15 Instrucciones Figuras continuación 7 A 8 B 9 C D ...
Page 23: ...0408 600 705 Rev AB DE www stryker com 23 Anleitung Abbildungen 1 3 2 4 2a 5 2b 6 ...
Page 24: ...DE 0408 600 705 Rev AB 24 www stryker com Anleitung Abbildungen Fortsetzung 7 A 8 B 9 C D ...
Page 32: ...FR 0408 600 705 Rev AB 32 www stryker com Mode d emploi Figures 1 3 2 4 2a 5 2b 6 ...
Page 33: ...0408 600 705 Rev AB FR www stryker com 33 Mode d emploi Figures suite 7 A 8 B 9 C D ...
Page 41: ...0408 600 705 Rev AB IT www stryker com 41 Istruzioni Figure 1 3 2 4 2a 5 2b 6 ...
Page 42: ...IT 0408 600 705 Rev AB 42 www stryker com Istruzioni Figure continua 7 A 8 B 9 C D ...
Page 50: ...NL 0408 600 705 Rev AB 50 www stryker com Instructies Afbeeldingen 1 3 2 4 2a 5 2b 6 ...
Page 51: ...0408 600 705 Rev AB NL www stryker com 51 Instructies Afbeeldingen vervolg 7 A 8 B 9 C D ...
Page 59: ...0408 600 705 Rev AB SV www stryker com 59 Anvisningar Figurer 1 3 2 4 2a 5 2b 6 ...
Page 60: ...SV 0408 600 705 Rev AB 60 www stryker com Anvisningar Figurer forts 7 A 8 B 9 C D ...
Page 68: ...DA 0408 600 705 Rev AB 68 www stryker com Anvisninger Figurer 1 3 2 4 2a 5 2b 6 ...
Page 69: ...0408 600 705 Rev AB DA www stryker com 69 Anvisninger Figurer fortsat 7 A 8 B 9 C D ...
Page 77: ...0408 600 705 Rev AB FI www stryker com 77 Ohjeet Kuvat 1 3 2 4 2a 5 2b 6 ...
Page 78: ...FI 0408 600 705 Rev AB 78 www stryker com Ohjeet Kuvat jatkoa 7 A 8 B 9 C D ...
Page 86: ...PT 0408 600 705 Rev AB 86 www stryker com Instruções Figuras 1 3 2 4 2a 5 2b 6 ...
Page 87: ...0408 600 705 Rev AB PT www stryker com 87 Instruções Figuras continuação 7 A 8 B 9 C D ...
Page 95: ...0408 600 705 Rev AB NO www stryker com 95 Instruksjoner Figurer 1 3 2 4 2a 5 2b 6 ...
Page 96: ...NO 0408 600 705 Rev AB 96 www stryker com Instruksjoner Figurer fortsettelse 7 A 8 B 9 C D ...
Page 104: ...PL 0408 600 705 Rev AB 104 www stryker com Instrukcje Rysunki 1 3 2 4 2a 5 2b 6 ...
Page 105: ...0408 600 705 Rev AB PL www stryker com 105 Instrukcje Rysunki ciąg dalszy 7 A 8 B 9 C D ...
Page 113: ...0408 600 705 Rev AB EL www stryker com 113 Οδηγίες Εικόνες 1 3 2 4 2a 5 2b 6 ...
Page 114: ...EL 0408 600 705 Rev AB 114 www stryker com Οδηγίες Εικόνες συνέχεια 7 A 8 B 9 C D ...
Page 119: ......