Care and Maintenance
37
EN
• type and scope of the performed services,
• date of service,
• and service company with signature.
9.8
Replacing the Fuse
The fuse may be defective and is in need of replacement if
• the device is connected with the power supply, the green power LED of the
pump unit is not illuminated after pressing the ON/OFF switch (
I
) and the
device does not initialize or perform a device test,
• the device does not function.
Check to make sure
• the main power supply cable is properly connected to the non-heating device
plug and a grounded safety socket,
• whether the house power supply fuse is functioning.
Unplug the power cable from the device before checking the fuse.
9.8.1 How to Replace the Fuse
The device does not have to be opened to replace the fuse.
1. Switch device off.
2. Disconnect device from power supply.
3. Remove power cable from device plug
(2)
.
4. The fuse holder
(3)
is located right next to the non-heating device plug.
Remove fuse holder as depicted in Fig. 9-1 Opening the fuse holder, page 37.
5. Check the condition and values of the installed / to be installed fuse (4)
according to chapter 12 Technical Data, page 43.
6. Follow these steps in reverse for assembly. The fuse holder has to clearly snap
into place when installing and has to be aligned flush with the non-heating
device plug.
7. Use the mains power cable to reconnect the shockproof socket with the
non-heating device plug.
DANGER
(1)
(2)
(4)
(3)
Fig. 9-1
Opening the fuse holder
(1)
Screwdriver
(2)
Non-heating device plug
(3)
Fuse holder
(4)
Fuse
Summary of Contents for FLUID SAFE
Page 1: ...EN ES FR DE Manual Manual Manuel Handbuch 2016 09 www stryker com 1000 400 948 Rev F ...
Page 6: ......
Page 8: ......
Page 55: ...Appendix 49 EN 15 Appendix 15 1 Test Record Date Result Comment Signature ...
Page 58: ......
Page 60: ......
Page 109: ...Apéndice 49 ES 15 Apéndice 15 1 Registro de test Fecha Resultado Anotaciones Firma ...
Page 114: ......
Page 162: ...Annexe 48 FR 15 Annexe 15 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 166: ......
Page 168: ......
Page 216: ...Anhang 48 DE 15 Anhang 15 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Page 220: ......
Page 221: ......