Operating the Device
28
EN
5.13 Launching/Configuring the User Menu
Device parameters are changed and displayed with the
user menu
.
• Hold down the menu key
(MENU)
and turn on the device using the ON/OFF key.
• Keep the
(MENU)
key depressed until the LED test is finished. The nominal flow
display will depict a
P
after the LED test and emit an acoustic signal. The device
has now entered the first menu level of the menu mode.
• The first menu level offers a choice between
User menu
and
Service menu
. The
nominal pressure display depicts
P
for User menu or
L
for Service menu
(accessible only to service technicians).
•
Switch
from one to the other with the
▲
/
▼
keys.
•
Confirm
with the
(MENU)
key.
•
Select
the menu level or the value to be adjusted using the
▲
/
▼
keys.
• The values to be modified are
displayed
in the nominal flow display.
• Adjusted values are
saved
with the
(MENU)
key (use like ENTER key).
•
Exit
the menu or menu level without saving with the Start/Stop key
functioning as the ESC key.
• To reenter OP mode, turn the device off and on again.
The following user menu levels and settings are available:
Parameter
Description
Factory
Settings
P
User menu (accessible to the user)
L
Service menu (accessible only to service personnel)
P01
Volume level of pump unit alarm.
Two possible settings (1=muted, 2=loud).
2
P02
Setting, if balancing unit connected or not.
0 = not connected
1 = connected
first saved, then turn on again.
After saving the next menu points are not longer accessible.
1
P03
Setting for the deficit threshold - acoustic warning levels in the balancing unit
Setting range 0-2000 ml.
When pressing the key briefly -> in increments of 10
When pressing the key longer -> in increments of 100
Additional depiction in the deficit display
750 ml
P04
Volume of the acoustic warning in the balancing unit
Two possible settings (1=muted, 2=loud).
Additional depiction in the deficit display.
2
P05
Check of the mass gauge of the balancing unit
Display in windows for deficit (0) – and nominal flow (0.00).
Nominal
Pressure
80 mmHg
Nominal
Flow
150 ml/min
Summary of Contents for FLUID SAFE
Page 1: ...EN ES FR DE Manual Manual Manuel Handbuch 2016 09 www stryker com 1000 400 948 Rev F ...
Page 6: ......
Page 8: ......
Page 55: ...Appendix 49 EN 15 Appendix 15 1 Test Record Date Result Comment Signature ...
Page 58: ......
Page 60: ......
Page 109: ...Apéndice 49 ES 15 Apéndice 15 1 Registro de test Fecha Resultado Anotaciones Firma ...
Page 114: ......
Page 162: ...Annexe 48 FR 15 Annexe 15 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 166: ......
Page 168: ......
Page 216: ...Anhang 48 DE 15 Anhang 15 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Page 220: ......
Page 221: ......