Error and Warning Messages
45
EN
E01
• Service icon is lit
• Continuous acoustic alarm
• E01 flashes in the actual pressure
display.
• Pressure measuring failure
/malfunction.
• Pressure sensor 1 and/or 2 defective.
• Replace the device when error occurs
permanently.
E02
• Service icon is lit
• Continuous acoustic alarm
•
E02
flashes in the actual pressure
display
• Failure of the weighing unit / display
unit Data lines or weighing unit
defective.
• Replace device.
• Pump without balancing still to be
used for diagnosis.
E50
• Service icon is lit
• Acoustic alarm, 1x
•
E50
flashes in the actual pressure
display
• Communication error
• Malfunctioning communication
between modules.
• Replace device.
E51
• Service icon is lit
• Acoustic alarm, 2x
•
E51
flashes in the actual pressure
display
• Communication error
• Electronics error
• Replace device.
E59
• Acoustic alarm, 10x
•
E59
flashes in the actual pressure
display
• Safe status, only M key is active
• Self-test calibration error
• Initial calibration not executed or
interrupted.
•New calibration required.
No display in the display unit.
Scale connection not activated in the
user menu.
Activate the scale connection in the
user menu item
P02
.
Summary of Contents for FLUID SAFE
Page 1: ...EN ES FR DE Manual Manual Manuel Handbuch 2016 09 www stryker com 1000 400 948 Rev F ...
Page 6: ......
Page 8: ......
Page 55: ...Appendix 49 EN 15 Appendix 15 1 Test Record Date Result Comment Signature ...
Page 58: ......
Page 60: ......
Page 109: ...Apéndice 49 ES 15 Apéndice 15 1 Registro de test Fecha Resultado Anotaciones Firma ...
Page 114: ......
Page 162: ...Annexe 48 FR 15 Annexe 15 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 166: ......
Page 168: ......
Page 216: ...Anhang 48 DE 15 Anhang 15 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Page 220: ......
Page 221: ......