Operating the Device
21
EN
1. Remove the transport safeguard from the pressure chamber.
2.
A
Carefully insert the unpressurized pressure chamber
(6)
into the notch of
the pump head (nose
(4)
up). Listen for the click of the locking pressure
chamber. A properly inserted pressure chamber is flush with the pump head.
3. Place roller tube
(1)
around the roller wheel.
4.
B
Adjust the tube fastener
(2)
by tightening the tube at the upper end of the
pump head. Make sure the tube fastener is firmly snapped into place!
5. Connect an instrument and the corresponding bags.
Suspension of the fluid bags
The fluid bags may only be hanged on the hooks intended for fluid bags. We
strongly urge using only fluid containers made of flexible materials (plastic bags).
The max. allowed dimensions for the bags are:
Width: 300 mm/11.81 inches
Height: 440 mm/17.32 inches
The device is only intended for use with flexible fluid containers. Do not use glass
containers as they might break. Fluid cannot flow quickly enough due to the
vacuum being generated inside of the bottle. Risk of implosion.
Take the cap off the instrument tube. Connect the instrument tube
(5)
to the
instrument.
Connecting the irrigation fluid bags
Connect the clamps to the irrigation tube. The irrigation tube is prepared for the
removal of irrigation fluid from two fluid bags. When connecting or removing the
tube to or from the irrigation fluid container, grasp the insertion spike at the
provided handle. Observe sterile conditions when inserting the insertion spike
into the container.
The surgeon has to select an irrigation fluid suitable for the application and
medical procedure.
The main connections of the tube and device items are shown in Fig. 5-7.
DANGER
(1)
(5)
(6)
(7)
(8)
(9)
(10)
(2)
(3)
(4)
Fig. 5-7
Diagram of the tube
connections
(1)
Irrigation fluid bag
(2)
Tap spike
(3)
Irrigation tube
(4)
Pump head
(5)
Instrument (Resectoscope)
(6)
Apron
(7)
Suctioning
(8)
Connection tube for container
(9)
Container
(10)
Outflow tube
Summary of Contents for FLUID SAFE
Page 1: ...EN ES FR DE Manual Manual Manuel Handbuch 2016 09 www stryker com 1000 400 948 Rev F ...
Page 6: ......
Page 8: ......
Page 55: ...Appendix 49 EN 15 Appendix 15 1 Test Record Date Result Comment Signature ...
Page 58: ......
Page 60: ......
Page 109: ...Apéndice 49 ES 15 Apéndice 15 1 Registro de test Fecha Resultado Anotaciones Firma ...
Page 114: ......
Page 162: ...Annexe 48 FR 15 Annexe 15 1 Procès verbal de tests Date Résultat Remarque Signature ...
Page 166: ......
Page 168: ......
Page 216: ...Anhang 48 DE 15 Anhang 15 1 Testprotokoll Datum Ergebnis Bemerkung Unterschrift ...
Page 220: ......
Page 221: ......