14 January 2021, SmartScanIS User’s Guide for Union Pacific, EUD-2016026-00 Rev.2
6
Mate the Molex housing (on the end of the cable of the probe) to the factory-wired
Molex socket on the end of the RF-filter assembly.
7
Using a wire stripper, remove 1/4 inches (6.4 millimeters) of insulation from the ends of
the four wires coming from the other end of the RF-filter assembly.
On the Surge Protection Subassembly (in the SmartScanIS enclosure), the two UTBs
labeled
TempProbe
are for the shielded temperature probe. Each UTB has four rows
of connectors. As shown below, the wires from the shielded temperature probe are
terminated at the third row and fourth row of connectors from the top. The other
connectors on these UTBs are prewired at the factory and need no further wiring.
8
As shown above, terminate the wires from the end of the RF-filter assembly.
9
Mount the two filters of the RF-filter assembly onto the inside of the bungalow.
Black
Red
White
Shield
Black
Red
White
White
Shield
Shield
Summary of Contents for SmartScanIS
Page 1: ...EUD 2016026 00 Rev 2 SmartScanIS User s Guide for Union Pacific 14 January 2021...
Page 82: ...82 14 January 2021 SmartScanIS User s Guide for Union Pacific EUD 2016026 00 Rev 2...
Page 90: ...90 14 January 2021 SmartScanIS User s Guide for Union Pacific EUD 2016026 00 Rev 2...
Page 130: ...130 14 January 2021 SmartScanIS User s Guide for Union Pacific EUD 2016026 00 Rev 2...
Page 258: ...258 14 January 2021 SmartScanIS User s Guide for Union Pacific EUD 2016026 00 Rev 2...
Page 376: ...376 14 January 2021 SmartScanIS User s Guide for Union Pacific EUD 2016026 00 Rev 2...
Page 380: ...380 14 January 2021 SmartScanIS User s Guide for Union Pacific EUD 2016026 00 Rev 2...