1-28 Octave
Analysis
SR780 Network Signal Analyzer
8. Press [Display Setup]
Select the Display Setup menu.
Press
<Measurement>
Change the Measurement of both displays.
Remember, in 1 channel analysis, both displays
measure the same input.
Use the knob to select (Oct ch2) and press
[Enter].
Both displays now show the filter output on Ch2.
Press
<Measurement>
Change the Measurement again.
Use the knob to select (Oct ch1) and press
[Enter].
Show the source output on Ch1 on both displays.
9. Press [Input]
Select the Input menu.
Press
<More>
Show More Input.
Press <Ch1 A-Wt Filter>
Choose the Ch1 Input A-Weighting filter.
Use the knob to select (On) and press [Enter].
Turn the A-Wt filter On. The Ch1 Awt indicator at
the top of the screen is highlighted.
The hardware A-Wt filter conforms to the ANSI
standard and is commonly used in sound
measurements. This filter attenuates high and low
frequencies according to how people hear and
perceive sound.
Do not use this filter unless A-Weighting is
required since the signal is greatly altered by this
filter.
Press [Link] [Auto Scale A]
The A-Weighted spectrum is a bandpass centered
around 2 kHz.
This concludes this measurement example. You
should have a feeling for Octave measurements and
how they are setup.
Summary of Contents for SR780
Page 4: ...ii SR780 Network Signal Analyzer ...
Page 10: ...viii SR780 Network Signal Analyzer ...
Page 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Page 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Page 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Page 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Page 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Page 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Page 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Page 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Page 232: ......
Page 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Page 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Page 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Page 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Page 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Page 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Page 478: ......