4-56 Sine Source Menu
SR780 Network Signal Analyzer
Offset
Set the DC Offset of the Sine Source [-5V..5V]. The offset resolution is 0.1 mV. The sine
output is the sum of two tones (sine waves) and the offset (constant). Note that the sum of
the amplitudes of Tone 1, Tone 2 and the absolute value of the offset cannot exceed 5 V.
Also, using large offsets with small tone amplitudes will degrade the distortion
performance of the sine source.
Command: SOFF (?) {x}
Tone 2
Display the Tone 2 Settings menu. Press <Return> for the main [Source] menu.
Frequency 2
Set the Frequency of Tone 2. The sine output is the sum of two tones (sine waves). When
the knob is used to adjust the frequency, the resolution is equal to the Linewidth of the
active display FFT Span. The knob always sets the frequency to an exact multiple of the
Linewidth. For octave analysis, the Linewidth of the most recent FFT Spans are used
(with an FFT Base of 100.0 kHz).
The keypad allows arbitrary frequencies to be entered. Remember, the output is periodic
over the FFT time record only if the frequency is an exact multiple of the Linewidth.
Source Trigger will not result in a stable phase for non-periodic frequencies.
The sine source is not triggered. The output is always continuous.
Command: S2FR (?) {f}
Amplitude 2
Set the Amplitude of Tone 2 [0..5V]. The amplitude resolution is 0.1 mV. The sine output
is the sum of two tones (sine waves) and the DC offset (constant). To generate a single
tone, set the amplitude of one of the tones to zero. Note that the sum of the amplitudes of
Tone 1, Tone 2 and the absolute value of the offset cannot exceed 5 V.
Command: S2AM (?) {x}
Summary of Contents for SR780
Page 4: ...ii SR780 Network Signal Analyzer ...
Page 10: ...viii SR780 Network Signal Analyzer ...
Page 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Page 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Page 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Page 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Page 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Page 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Page 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Page 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Page 232: ......
Page 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Page 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Page 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Page 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Page 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Page 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Page 478: ......