1-44 Swept Sine Measurement
SR780 Network Signal Analyzer
Press <# of Points>
Adjust the Number Of Points in the sweep.
Press [1] [0] [2] [4] [Enter].
Enter 1024 points.
5. Press [Start/Reset]
Reset and start the sweep. The source is a sine wave
whose frequency sweeps from 900 Hz to 1.1 kHz
stopping at 1024 discrete frequencies. At each
frequency point, the inputs are measured and
displayed.
Wait for the sweep to complete at least once.
The small triangular marker moving across the
bottom of the graphs indicates the position of the
sweep in progress.
Press [Auto Scale A] and [Auto Scale B]
Scale the two displays. The top display is the filter
input (source output) and should be fairly flat. The
bottom display is the filter output and show the
filter notch at 1 kHz.
6. Press [Display Setup]
Select the Display Setup menu.
Press
<Measurement>
Change the Measurement of DisplayA (active
display).
Select (Transfer Function) with the knob and
press [Enter].
Choose Transfer Function (filter output divided by
filter input).
Press
<Units>
Transfer function is unitless. Change the display
units.
Select (dB) with the knob and press [Enter].
Choose dB.
Press [Display Options]
Select the Display Options menu.
Press
<Format>
Change the Display Format.
Select (Single) with the knob and press [Enter].
Choose a Single Display with DisplayA (Transfer
Function) active.
Press [Auto Scale A]
Scale the display to show the Transfer Function.
7. Press [Marker]
Select the Marker menu.
Press
<Width>
Change the Marker Region width.
Select (Normal) with the knob and press [Enter].
Choose Normal Width (1/2 division).
Press
<Seeks>
Change the Marker Seeks function.
Summary of Contents for SR780
Page 4: ...ii SR780 Network Signal Analyzer ...
Page 10: ...viii SR780 Network Signal Analyzer ...
Page 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Page 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Page 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Page 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Page 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Page 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Page 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Page 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Page 232: ......
Page 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Page 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Page 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Page 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Page 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Page 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Page 478: ......