5-60 Average
Commands
SR780 Network Signal Analyzer
Average Commands (Swept Sine)
SSTM (?) d {, x}
The SSTM command sets (queries) the Swept Sine Settle Time for display d. The
parameter x is a settle time from 0.008 to 1000 seconds. The set command
requires d=2 (both displays).
This command is valid only when the Measurement Group is Swept Sine. The set
command requires a display to be Live.
SSCY (?) d {, i}
The SSCY command sets (queries) the Swept Sine Settle Cycles for display d.
The parameter i is a number of cycles from 1 to 32767 seconds. The set
command requires d=2 (both displays).
This command is valid only when the Measurement Group is Swept Sine. The set
command requires a display to be Live.
SITM (?) d {, x}
The SITM command sets (queries) the Swept Sine Integration Time for display
d. The parameter x is an integration time from 0.016 to 1000 seconds. The set
command requires d=2 (both displays).
This command is valid only when the Measurement Group is Swept Sine. The set
command requires a display to be Live.
SICY (?) d {, i}
The SICY command sets (queries) the Swept Sine Integration Cycles for display
d. The parameter i is a number of cycles from 1 to 32767 seconds. The set
command requires d=2 (both displays).
This command is valid only when the Measurement Group is Swept Sine. The set
command requires a display to be Live.
Summary of Contents for SR780
Page 4: ...ii SR780 Network Signal Analyzer ...
Page 10: ...viii SR780 Network Signal Analyzer ...
Page 18: ...1 2 Getting Started SR780 Network Signal Analyzer ...
Page 39: ...Triggering and the Time Record 1 23 SR780 Network Signal Analyzer ...
Page 40: ...1 24 Triggering and the Time Record SR780 Network Signal Analyzer ...
Page 58: ...1 42 Waterfall Display SR780 Network Signal Analyzer ...
Page 70: ...1 54 Saving and Recalling SR780 Network Signal Analyzer ...
Page 167: ...Status Indicators 3 21 SR780 Network Signal Analyzer ...
Page 188: ...4 6 Menus SR780 Network Signal Analyzer ...
Page 193: ...FFT Frequency Menu 4 11 SR780 Network Signal Analyzer Command FEND d f ...
Page 232: ......
Page 236: ...4 54 Source Menu SR780 Network Signal Analyzer Command STYP i ...
Page 266: ...4 84 Playback Input Menu SR780 Network Signal Analyzer ...
Page 272: ...4 90 Trigger Menu SR780 Network Signal Analyzer ...
Page 288: ...4 106 Swept Sine Average Menu SR780 Network Signal Analyzer ...
Page 326: ...144 Exceedance Statistics Menu SR780 Network Signal Analyzer ...
Page 352: ...4 170 Output Menu SR780 Network Signal Analyzer ...
Page 478: ......